Description
Taylor K-1831
ChemWorld is a distributor of Taylor Technologies Test Kits and Reagents. Order directly online and save today.2 STANDARD COMPARATOR TEST
ORTHOPHOSPHATE (HIGH) 2 STANDARD
9029 (5, 10 ppm)
9026 (20, 40 ppm)
9028 (20, 60 ppm)
9025 (30, 60 ppm)
t9030 (40, 80 ppm)
*Kit may only include one.
REAGENTS: R-0601 Molybdate Reagent* R-0602P Stannous Chloride Powder APPARATUS: 2 x 2269 Caps, 18 mm Dispenser, white 1 x 4027 Funnel, 58 mm, plastic 1 x 5210 Instruction 1 x 6003 Brush, Test Tube 1 x 6009 Filter Paper, #610, 9.0 cm, 100/box 1 x 9021 Test Tube, Mixing, Calibrated 5, 10, 14, 15, 17.5 mL, glass w/stopper
1. Filter water to be tested to clarify.
2. Rinse and fill mixing test tube (#9021) to 5 mL mark with filtered sample.
3. Cap R-0601 Molybdate Reagent* with 18 mm dispenser cap (#2269). Add R-0601 Molybdate Reagent* to 15 mL mark. Stopper and mix. Remove stopper.
4. Add 2 level dippers R-0602P Stannous Chloride Powder. Stopper and mix.
5. Wipe dry and place in comparator. WAIT 1 MINUTE. Read before 10 minutes.
6. Match color in mixing test tube with a color standard. Record as parts per million (ppm) orthophosphate.
*WARNING: Molybdate Reagent, R-0601, contains 8.1% sulfuric acid, a corrosive acid.